THE EFFECT OF CONCENTRATION AND RINSE ON THE RETENTION OF LIQUID IODIUM-131 RADIOACTIVITY IN THE SYRINGE IN THE PREPARATION OF RADIOPHARMACIES
DOI:
https://doi.org/10.36387/jiis.v11i2.3047Keywords:
Iodine-131, radiopharmaceutical, radioactivity retention, radioactive contamination, radiation exposure, nuclear medicineAbstract
Radioactivity retention in syringes during the preparation of liquid Iodine-131 radiopharmaceuticals may potentially reduce the prepared activity, cause radioactive contamination, decrease imaging quality, increase radioactive waste, and contribute to unwanted radiation exposure. This study aimed to evaluate the effects of concentration and rinsing on radioactivity retention of liquid Iodine-131 in syringes. An experimental study was conducted using liquid Iodine-131 at concentrations of 5, 10, and 15 mCi/mL. Initial and residual activities in the syringe were measured using a dose calibrator, while radiation exposure rate was measured using a gamma survey meter. The rinsing test was performed at a concentration of 15 mCi/mL using water for injection (WFI) and 0.9% NaCl as rinsing solutions in three sequential rinsing steps. Radioactivity retention increased with increasing concentration, with values of 12.90 ± 0.63%, 17.45 ± 0.68%, and 20.97 ± 0.74% at concentrations of 5, 10, and 15 mCi/mL, respectively. Radiation exposure rate also increased significantly. Rinsing significantly reduced radioactivity retention, while the type of rinsing solution used showed no significant difference. Concentration and rinsing affected radioactivity retention of liquid Iodine-131 in syringes
References
Badan Pengawas Obat dan Makanan. (2024). Peraturan Badan Pengawas Obat dan Makanan Nomor 7 Tahun 2024 tentang Standar Cara Pembuatan Obat yang Baik.
Bergeron, D. E., & Cessna, J. T. (2018). An update on ‘dose calibrator’ settings for nuclides used in nuclear medicine. Nuclear Medicine Communications, 39(6), 500–504. https://doi.org/10.1097/MNM.0000000000000833
Cambruzzi, M., & Macfarlane, P. (2021). Variation in syringes and needles dead space compared to the International Organization for Standardization standard 7886-1:2018. Veterinary Anaesthesia and Analgesia, 48(4), 532–536. https://doi.org/10.1016/j.vaa.2021.01.008
Galbraith, W., Chen, X., Talley, K., & Grantham, V. (2015). Assessment of 99mTc-Succimer Residual Activity Using Inert Nonreactive Syringes. Journal of Nuclear Medicine Technology, 43(1), 61–63. https://doi.org/10.2967/jnmt.114.147983
Gillings, N., Hjelstuen, O., Ballinger, J., Behe, M., Decristoforo, C., Elsinga, P., Ferrari, V., Peitl, P. K., Koziorowski, J., Laverman, P., Mindt, T. L., Neels, O., Ocak, M., Patt, M., & Todde, S. (2021). Guideline on current good radiopharmacy practice (cGRPP) for the small-scale preparation of radiopharmaceuticals. EJNMMI Radiopharmacy and Chemistry, 6(1), 8. https://doi.org/10.1186/s41181-021-00123-2
Haugen, B. R., Alexander, E. K., Bible, K. C., Doherty, G. M., Mandel, S. J., Nikiforov, Y. E., Pacini, F., Randolph, G. W., Sawka, A. M., Schlumberger, M., Schuff, K. G., Sherman, S. I., Sosa, J. A., Steward, D. L., Tuttle, R. M., & Wartofsky, L. (2016). 2015 American Thyroid Association Management Guidelines for Adult Patients with Thyroid Nodules and Differentiated Thyroid Cancer: The American Thyroid Association Guidelines Task Force on Thyroid Nodules and Differentiated Thyroid Cancer. Thyroid, 26(1), 1–133. https://doi.org/10.1089/thy.2015.0020
Jirsa, L., Varga, F., & Quinn, A. (2017). Identification of thyroid gland activity in radioiodine therapy. Informatics in Medicine Unlocked, 7, 23–33. https://doi.org/10.1016/j.imu.2017.02.004
Keskintepe, D., & Özer, A. Y. (2005). Residual radioactivity in the syringes used in injection of radiopharmaceuticals and factors affecting the amount of residual radioactivity. Fabad Journal of Pharmaceutical Sciences, 30(4), 176–180.
Kesten, J. M., Ayres, R., Neale, J., Clark, J., Vickerman, P., Hickman, M., & Redwood, S. (2017). Acceptability of low dead space syringes and implications for their introduction: A qualitative study in the West of England. International Journal of Drug Policy, 39, 99–108. https://doi.org/10.1016/j.drugpo.2016.09.005
Kvaternik, H., Gatterer, J., Plhak, E., Schwarzgruber, J. F., & Aigner, R. M. (2020). Systematic Assessment of the Adsorption of99m Tc-Radiopharmaceuticals onto Plastic Syringes. Journal of Nuclear Medicine Technology, 48(2), 168–173. https://doi.org/10.2967/jnmt.119.235432
Ladrière, T., Aussage, L., & Vigne, J. (2024). Container-content interactions with radiopharmaceuticals: Seeing is believing. European Journal of Pharmaceutics and Biopharmaceutics, 196, 114200. https://doi.org/10.1016/j.ejpb.2024.114200
Mushtaq, A., Ur Rehman, T., Safdar Mansur, M., & Jehangir, M. (2008). Adsorption of 99mTc-Radiopharmaceuticals onto Injection Vials and Syringes. Journal of Nuclear Medicine Technology, 36(2), 91–94. https://doi.org/10.2967/jnmt.107.048561
Nguyen, N. C., Anigati, E. M., Desai, N. B., & Öz, O. K. (2024). Radioactive Iodine Therapy in Differentiated Thyroid Cancer: An Update on Dose Recommendations and Risk of Secondary Primary Malignancies. Seminars in Nuclear Medicine, 54(4), 488–496. https://doi.org/10.1053/j.semnuclmed.2024.05.002
Paquet, F., Etherington, G., Bailey, M. R., Leggett, R. W., Lipsztein, J., Bolch, W., Eckerman, K. F., & Harrison, J. D. (2015). ICRP Publication 130: Occupational Intakes of Radionuclides: Part 1. Annals of the ICRP, 44(2), 5–188. https://doi.org/10.1177/0146645315577539
Stavrou, P. Z., Papachristou, M., Persakis, E., Kouvelis, K., & Datseris, I. E. (2016). Residual activities of 99mTc-labelled radiopharmaceuticals in routine nuclear medicine practice. Nuclear Medicine Communications, 37(6), 658–663. https://doi.org/10.1097/MNM.0000000000000487
Swanson, T. N., Troung, D. T., Paulsen, A., Hruska, C. B., & O’Connor, M. K. (2013). Adsorption of 99mTc-Sestamibi onto Plastic Syringes: Evaluation of Factors Affecting the Degree of Adsorption and Their Impact on Clinical Studies. Journal of Nuclear Medicine Technology, 41(4), 247–252. https://doi.org/10.2967/jnmt.113.132159
Wadsley, J., Armstrong, N., Bassett-Smith, V., Beasley, M., Chandler, R., Cluny, L., Craig, A. J., Farnell, K., Garcez, K., Garnham, N., Graham, K., Hallam, A., Hill, S., Hobrough, H., McKiddie, F., & Strachan, M. W. J. (2023). Patient Preparation and Radiation Protection Guidance for Adult Patients Undergoing Radioiodine Treatment for Thyroid Cancer in the UK. Clinical Oncology, 35(1), 42–56. https://doi.org/10.1016/j.clon.2022.07.002